DM DOC 040 Rev 1.0
Data derived from 47 cases of paced ECGs.
PACING
STATEMENT
SENS
SPEC
PPV
NPV
PREV
Atrial Pacing
100%
97.3%
90.9%
100%
10/47
Demand Atrial
Pacing
100%
100%
100%
100%
5/47
Ventricular
pacing
90.9%
100%
100%
97.3%
11/47
A-V Sequential
Pacemaker
90.9%
100%
100%
97.3%
11/47
Demand pacing
100%
97.3%
90.9%
100%
10/47
Type C statements in adults derived from the Glasgow 1000 ECG database less
the 73 childrens’ ECGs.
ECG FINDINGS
SENS
SPEC
PPV
NPV
PREV
LAD
100%
100%
100%
100%
65/927
Leftward axis
100%
100%
100%
100%
59/927
RAD
100%
100%
100%
100%
15/927
Severe RAD
100%
100%
100%
100%
1/927
Rightward axis
100%
100%
100%
100%
12/927
Non specific ST
T changes
100%
99.8%
97.3%
100%
71/927
rSr' – probable
normal variant
100%
99.9%
94.4%
100%
17/927
Poor R wave
progression
92.8%
99.9%
96.3%
99.8%
26/927
LAD = Left axis deviation
RAD = Right axis deviation
© 2013 DanMedical Ltd