![DanMedical D-MAS HyperSat 2540 Instructions For Use Manual Download Page 71](http://html1.mh-extra.com/html/danmedical/d-mas-hypersat-2540/d-mas-hypersat-2540_instructions-for-use-manual_3416622071.webp)
DM DOC 040 Rev 1.0
Resting ECG Screen
WARNING!
Defibrillation and Electrosurgery: Do not touch the patient, or table, or instruments,
during defibrillation.
After defibrillation, the screen display recovers within 10 seconds if the correct electrodes are
used and applied in accordance with the manufacturers instructions.
ECG cables can be damaged when connected to a patient during defibrillation. Check cables for
functionality before using them again.
When using electrosurgical (ES) equipment, never place ECG electrodes near to the grounding
plate of the ES device, as this can cause a lot of interference on the ECG signal.
When you are connecting the electrodes or the patient cable, make sure that the connectors
never come into contact with other conductive parts, or with earth. In particular, make sure that
all of the ECG electrodes are attached to the patient, to prevent them from contacting conductive
parts or earth.
Preparing the Patient
Proper patient preparation and correct
electrode placement is important.
Perform the following steps to prepare the
patient.
Relax the patient
Remove hair at electrode site
Gently abrade the skin
Place the electrodes on the
patient
Attach each patient lead wire to
the correct electrode
D-MAS Remote ECG Socket
12 Lead ECG electrode placement
Place 10 skin electrodes on the
patient.
Connect the electrode clips to the
patient skin electrodes
Connect the 12 lead ECG cable to
the D-MAS ECG socket.
Launch D-MAS software to view
ECG.
Limb lead electrode placement.
© 2013 DanMedical Ltd
71