DM DOC 040 Rev 1.0
MENU ITEMS
DESCRIPTION
SpO2 HIGH LIMIT
The SpO2 high alarm limit can be set anywhere between 2% and 99%, with
a 1% step size. In the “----” (off) setting, the alarm can be turned off
completely.
SpO2 LOW LIMIT
The SpO2 low alarm limit can be set anywhere between 1% and 99%, with
a 1% step size.
NOTE:
The low alarm limit always has to be set below the high alarm
setting. When the high alarm limit is set below the low alarm limit, the low
alarm limit will automatically adjust to the next setting below the newly
entered high alarm limit setting.
PULSE RATE
HIGH LIMIT (BPM)
The pulse rate high alarm limit can be set anywhere between 30 BPM and
240 BPM, with a 5 BPM step size.
PULSE RATE LOW
LIMIT (BPM)
The pulse rate low alarm limit can be set anywhere between 25 BPM and
235 BPM, with a 5 BPM step size.
NOTE:
The low alarm limit always has to be set below the high alarm
setting. When the high alarm limit is set below the low alarm limit, the low
alarm limit will automatically adjust to the next setting below the newly
entered high alarm limit setting.
SpMet HIGH LIMIT
The SpMet high alarm limit can be set anywhere between 1% to 100%.
Between 1% and 2%, the step increment is 0.1%. Between 2% and 100%,
the step increment is 0.5%.
SpMet LOW LIMIT
The SpMet low alarm limit can be set anywhere between 0.1% to 99.5%.
Between 0.1% and 2%, the step increment is 0.1%. Between 2% and
99.5%, the step increment is 0.5%. In the “----” (off) setting, the alarm can
be turned off completely.
NOTE:
The low alarm limit must always be below the high alarm limit. If the
high alarm limit is set below the low alarm limit, the low alarm limit will
automatically adjust to the next setting below the newly entered high alarm
limit.
SpCO HIGH LIMIT
The SpCO high alarm limit can be set anywhere between 1% and 100%,
with a 1% step size.
SpCO LOW LIMIT
The SpCO low alarm limit can be set anywhere between 1% and 99%, with
a 1% step size. In the “----” (off) setting, the alarm can be turned off
completely.
NOTE:
The low alarm limit must always be below the high alarm limit. If the
high alarm limit is set below the low alarm limit, the low alarm limit will
automatically adjust to the next setting below the newly entered high alarm
limit
SILENCE
This menu allows the user to set the alarm silence period. An alarm is
silenced by clicking the Alarm Silence button on screen.
The alarm silence
can be set for these durations:
30, 60, 90,120 SECONDS
. As an indicator
that the alarm system is silenced, the Alarm Status Indicator is shown as a
bell with a slash through it. A timer is shown next to the bell indicating the
remaining alarm silence duration.
NOTE:
The alarm silence period is reset to 120 seconds upon power cycle.