DM DOC 040 Rev 1.0
LNOP® SPECIALTY SENSORS
(LNOP sensors must be used in conjunction with Red PC cables)
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
M
Motion
No Motion
Motion
Saturation
Pulse Rate
LNOP
Newborn
Infant
3 - 10 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
10 - 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP
Newborn
Neonatal
< 3 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNOP Blue
2.5 - 30 kg
60 - 80% ±
4%
N/A
± 3 bpm
N/A
± 3%
± 3 bpm
70 - 100% ±
3.3%
N/A
± 3 bpm
N/A
± 3%
± 3 bpm
80 - 100% ±
3%
N/A
± 3 bpm
N/A
± 3%
± 3 bpm
LNCSTM REUSABLE SENSORS
(LNCS sensors must be used in conjunction with Red LNC or Rainbow patient cables)
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNCS DC-I
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS DC-IP 10 - 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS TC-I
> 30 kg
± 3.5%
N/A
± 3 bpm
N/A
± 3.5%
± 3 bpm
LNCS TF-I
> 30 kg
± 2%
N/A
± 3 bpm
N/A
± 2%
± 3 bpm
NOTE:
The LNCS TF-I and TC-I sensors were not validated under motion conditions.
© 2013 DanMedical Ltd