DM DOC 040 Rev 1.0
MESSAGE
POSSIBLE CAUSE(S)
RECOMMENDATION
LOW SpCO CONF
• SpCO measurement reading
is obscured.
• Ensure proper sensor
application.
• Check sensor to see if it is
working properly.
• If not, replace the sensor.
LOW SpMet CONF
• SpMet measurement reading
is obscured.
• Ensure proper sensor
application. Check sensor to
see if it is working properly.
• If not, replace the sensor
NO SENSOR
• Sensor not fully inserted into
the connector.
• Maybe an incorrect sensor, or
a defective sensor or cable.
• Insert sensor into connector.
• Disconnect and reconnect
sensor.
• Refer to the instructions for
the sensor being used.
NO CABLE
• Cable not attached or not
fully inserted into the
connector.
• Disconnect and reconnect
cable into connector
• Unit is searching for patient’s
pulse.
• Disconnect and reconnect the
sensor into the Patient Cable
Connector.
PULSE SEARCH
• Unit is searching for patient’s
pulse.
• If values are not displayed
within 30 seconds, disconnect
and reconnect sensor.
• If pulse search continues,
remove sensor and replace on
a better perfused site.
SENSOR CALIBRATING
• Unit is checking the sensor
for proper functioning and
performance.
• If values are not displayed
within 30 seconds, disconnect
and reconnect sensor.
• If values are still not
displayed, replace with a new
sensor.
SENSOR OFF
• Sensor off patient.
• Disconnect and reconnect
sensor.
• Reattach sensor.
SERVICE REQUIRED*
• Internal Failure
• Unit requires service.
UNRECOGNIZED CABLE
• Not a proper cable.
• Replace with a proper cable.
Refer to Section 8.
*The SERVICE REQUIRED message is a numeric error code.
Contact DanMedical for service.