DM DOC 040 Rev 1.0
Spirometer function test
1. With D-MAS Software running, select the Spirometry button
2. Connect the Spirometer to a spare USB port on D-MAS
3. Observe blue power LED on Spirometer body
4. Click 'Start' and observe audible beep from Spirometer
5. Secureley connect a calibrated 3Litre syringe and push the maximum amount of air
through the Spirometer.
6. Observe graphic loop and measurement display.
7. Repeat test 3 times for repeatabilty.
8. If any degradations in performance are observed, contact technical assistance or refer
the D-MAS to qualified technical personnel.
© 2013 DanMedical Ltd