3.5.1 Saving and Restoring PRNU and FPN Coefficients
Saving the Current PRNU Coefficients
Saving the Current FPN Coefficients
Pu rp ose:
Saves the cu rrent FPN coefficients. You can save u p to fou r sets of
p ixel coefficients
Syntax:
wfc
i
Syntax Elem ents:
i
FPN coefficients set to save.
1
= Coefficient set one
2
= Coefficient set tw o
3
= Coefficient set three
4
= Coefficient set fou r
Exam p le:
wfc 2
Loading a Saved Set of Coefficients
Pu rp ose:
Load s a saved set of p ixel coefficients. A factory calibrated set of
coefficients is available.
Syntax:
lpc
i
Syntax Elem ents:
i
FPN coefficients set to save.
0 =
Factory calibrated p ixel coefficients.
1
= Coefficient set one
2
= Coefficient set tw o
3
= Coefficient set three
4
= Coefficient set fou r
Exam p le:
lpc 0
Resetting the Current Pixel Coefficients
Pu rp ose:
Resets the cu rrent p ixel coefficients to zero. This com m and d oes
not reset saved coefficients.
Syntax:
rpc
Pu rp ose:
Saves the cu rrent PRN U coefficients. You can save u p to fou r sets
of p ixel coefficients
Syntax:
wpc
i
Syntax Elem ents:
i
PRN U coefficients set to save.
1
= Coefficient set one
2
= Coefficient set tw o
3
= Coefficient set three
4
= Coefficient set fou r
Exam p le:
wpc 2