Returning Calibration Results and Errors
Returning All Pixel Coefficients
Pu rp ose:
Retu rns all the cu rrent p ixel coefficients in the ord er FPN , PRN U,
FPN, PRNU… for the range specified by
x1
and
x2
. The cam era
also retu rns the p ixel nu m ber w ith every fifth coefficient.
Syntax:
dpc
x1 x2
Syntax Elem ents:
x1
Start p ixel to d isp lay in a range from
1
to
sensor pixel
count
.
x2
End p ixel to d isp lay in a range from
1
to
sensor pixel
count
.
N otes:
This fu nction retu rns all the cu rrent p ixel coefficients in the
order FPN, PRNU, FPN, PRNU… The camera also returns the
p ixel nu m ber w ith each coefficient.
Exam p le:
dpc 10 20
Returning FPN Coefficients
Pu rp ose:
Returns a pixel’s FPN coefficient value in DN (12 bit LSB)
Syntax:
gfc
i
Syntax Elem ents:
i
The p ixel nu m ber to read in a range from
1
to
sensor
pixel count
.
Exam p le:
gfc 10
Setting a Range of FPN Coefficients
Pu rp ose:
Sets a range of p ixel FPN coefficients.
Syntax
sfr
x x i
Syntax Elem ents:
x
The first p ixel nu m ber of the range.
x
The last p ixel nu m ber of the range.
i
Coefficient valu e in a range from 0-2048.
N otes:
The first p ixel of the range m u st be less than the last.
Exam p le:
sfr 1 100 80