![Cynosure Cynergy Technical Manual Download Page 94](http://html.mh-extra.com/html/cynosure/cynergy/cynergy_technical-manual_2704655094.webp)
92 of 112
850-1265-000, Rev. 4
Cynosure, Inc.
Replacing the Deionizer (DI) Cartridge
1.
Drain the water.
2.
Remove the right side panel as detailed on page 54.
3.
Remove the old deionizer, located on the frame assembly by removing the tubing at the
fittings, and then opening the clamps. Remove the old cartridge.
4.
Insert a new cartridge, and then attach the tubing. If the cartridge does not have fittings,
the fittings must be transferred. Clean the threads of any old Teflon tape first. Then
place approximately 1.5 turns of new Teflon tape on the male threads, turning by hand
until tight.
Calibrating the Dye Concentration Monitor
1.
Turn on the laser.
2.
Enter Testall Mode as detailed on page 19.
3.
Attach the red lead from the DVM to TP20. Attach the black lead from the DVM to
TP3 (ground).
4.
From Testall mode, press lowercase ‘l’ to increase or uppercase ‘L’ to decrease the
voltage until the DVM reads from 1.99V to 2.010V.
Calibrating the Dye Temperature
1.
Turn on the laser.
2.
Enter Testall Mode as detailed on page 19.
3.
Place thermocouple probe into the dye reservoir through the hole provided for the
charcoal filter.
4.
Press lowercase ‘y’ and then enter the temperature reading on the DVM.
Summary of Contents for Cynergy
Page 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Page 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Page 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Page 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Page 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Page 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Page 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Page 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 115: ...Tabbed Page Final Test Calibration ...
Page 116: ......
Page 118: ......
Page 120: ......
Page 148: ......
Page 149: ...Tabbed Page Drawing Sets ...
Page 150: ......
Page 172: ......
Page 173: ...Tabbed Page Schematics ...
Page 174: ......
Page 176: ...This page intentionally left blank ...
Page 178: ...This page intentionally left blank ...
Page 180: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 184: ...This page intentionally left blank ...
Page 186: ...This page intentionally left blank ...
Page 188: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...