Cynosure, Inc.
Replacing IGBT Triple Driver PCB, PDL
1.
Unplug the Fastons connecting the PCB to the triple driver PCB. Keep track of which
wire goes to which connector, e.g., Faston 1 goes to Faston 2, Faston 2 goes to Faston 1.
2.
Remove the mounting screws for the PCB and the screws that mount direct to the IGBT.
All screws should be accessible with a long #2 Phillips.
3.
Remove the optical connectors, keeping track of their locations on reassembly.
4.
Slide the PCB out the side and reverse the procedure to install a new PCB.
Replacing Simmer/Snubber PCB, PDL
1.
Unplug the AC.
2.
Remove the mounting screws for the PCB and the screws that mount direct to the IGBT.
All screws should be accessible with a long #2 Phillips.
3.
Slide the PCB out the side and reverse the procedure to install a new PCB.
Replacing the Trigger Transformers
The trigger transformers strike a rapid high voltage spike to initiate breakdown of the lamps to
start the low current simmer. Once triggered, the lamps break down noticeably and produce a
brief blue glow. This is an indication that the transformer is working.
To replace the trigger transformers, perform the following steps.
1.
Detach the AC input to the trigger transformer and all other electrical terminations.
2.
Unscrew the transformer from the resonator mount.
3.
Remove the old simmer trigger transformer.
4.
Attach the new simmer trigger transformer.
Replacing the Simmer Transformers
The AC provided to the simmer supplies passes through two transformers, one for the YAG
(135 VA) and one for the PDL (170 VA). These are both located on the lower chassis near the
isolation transformer. One reason to replace the simmer transformers is if the AC is not present
at the input to the simmer PCBs, yet the fuses are good.
To replace the simmer transformers, perform the following steps.
1.
Remove the side skins. Identify the transformer that needs replacement.
2.
Remove the damaged simmer transformer.
3.
Attach the new simmer transformer.
4.
Verify that the low current simmer is maintained.
Summary of Contents for Cynergy
Page 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Page 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Page 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Page 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Page 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Page 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Page 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Page 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 115: ...Tabbed Page Final Test Calibration ...
Page 116: ......
Page 118: ......
Page 120: ......
Page 148: ......
Page 149: ...Tabbed Page Drawing Sets ...
Page 150: ......
Page 172: ......
Page 173: ...Tabbed Page Schematics ...
Page 174: ......
Page 176: ...This page intentionally left blank ...
Page 178: ...This page intentionally left blank ...
Page 180: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 184: ...This page intentionally left blank ...
Page 186: ...This page intentionally left blank ...
Page 188: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...