14 of 112
850-1265-000, Rev. 4
Cynosure, Inc.
Electrical Hazard
The system requires 220 VAC ± 10%, 30 A, 50-60 Hz, single-phase electrical service to operate.
WARNING: Even when the laser is off and the AC line cord is disconnected, DC
voltages on various laser components, such as capacitors may exist. This can present a
potentially fatal electrical hazard during service procedures.
Take the following precautions to avoid electrical shock during servicing:
♦
Always turn the laser off and disconnect the AC line cord from the receptacle before
removing the protective housing of the laser system.
♦
With the laser off, allow the dump resistors to dissipate the energy in the pulse forming
network capacitors to a safe level, approximately 45 seconds. Monitor the voltage on the
capacitors with a DVM probe to ensure it is at a safe level.
IMPORTANT: The Cynergy laser system, the capacitor bank does not dump when
going between laser states (Ready to Standby) or between laser wavelengths. The
only time it dumps is when the user selects Utility. The capacitor bank voltage is
displayed on the screen as a quick reference, but the voltage on the lamps should
still be checked with a DVM prior to servicing.
♦
If it is necessary to test or adjust any electrical component while the system power is on,
be careful not to touch any electrical components with bare fingers; use appropriate
probes or insulated tools only.
♦
Become familiar with the electrical schematics and layout of the system before
attempting to service the laser.
If the AC must be connected during service routines, exercise caution around mains connected
components, such as power supply feeds, circuit breakers, key switches, etc. A remote interlock
fault condition will reduce high voltage electrical hazards to service personal. This fault
condition disables the high voltage power supply control and simmer circuits, as well as ensuring
that the dump relay is closed, grounding the potential of the pulse forming network. When
possible, remove the remote interlock plug during “live” service routines.
Grounding
Even when the laser is turned off and the AC line cord disconnected, high DC voltage levels may
remain. Before performing any procedure, use a shorting stick to ground all interior components.
The main capacitor bank stores large amounts of electrical energy. Measure the voltage on the
capacitor bank before attempting any service. Ground the capacitor bank with a shorting stick for
at least five seconds.
WARNING: Do not attempt to short directly (i.e., with a screwdriver) or a potentially fatal
electrical shock can occur.
Summary of Contents for Cynergy
Page 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Page 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Page 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Page 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Page 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Page 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Page 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Page 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 115: ...Tabbed Page Final Test Calibration ...
Page 116: ......
Page 118: ......
Page 120: ......
Page 148: ......
Page 149: ...Tabbed Page Drawing Sets ...
Page 150: ......
Page 172: ......
Page 173: ...Tabbed Page Schematics ...
Page 174: ......
Page 176: ...This page intentionally left blank ...
Page 178: ...This page intentionally left blank ...
Page 180: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 184: ...This page intentionally left blank ...
Page 186: ...This page intentionally left blank ...
Page 188: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...