56 of 112
850-1265-000, Rev. 4
Cynosure, Inc.
Resonator Performance
The graph below represents typical Cynergy laser performance—actual performance will vary.
However, there should be no “roll-over,” meaning the output having little or no increase despite
an increase in voltage. When “roll-over” occurs it can usually be attributed to poor alignment.
See the “System Check” or “Laser Cal Condition Table” on the following page that flags errors
once the FOM has been determined. A minimum FOM performance of > 2.5 J (for the PDL)
and of
> 30 (for the YAG) is required for operation.
Cynergy (Output vs. Voltage)
0
10
20
30
40
50
60
70
80
300
400
500
600
700
800
900
1000
1100
Start Cap Bank Volts (V)
R
eso
n
a
to
r O
u
t (J)
YAG, 5 ms, 1*5.0
DYE, 0.5 ms
Figure 22–Resonator Performance
A separate performance standard for each laser has been established for the resonator output. In
order to meet the system performance criteria, the resonator must normally generate
approximately 15-30 % more output to compensate for losses in coupling to the delivery system.
For the PDL laser, the resonator must be capable of generating a minimum of 10 joules/pulse @
1Hz or 10 Watts. This is based on performance criteria of 10 J/cm
2
with a 10-mm handpiece,
which equals ~8 joules/pulse out of the handpiece. Adding a nominal 20% for delivery system
losses, this means that the resonator must produce 10 joules/pulse.
For the Nd:YAG laser, the resonator must be capable of generating a minimum of 78.5
joules/pulse @ 1Hz or 78.5 Watts. This is based is based on performance criteria of 80 J/cm
2
with a 10-mm handpiece, which equals 62.8 joules/pulse out of the handpiece. Adding a nominal
20% for delivery system losses, this means that the resonator must produce 78.5 joules/pulse.
If transmission is accepted as low as 52%, the laser heads must be capable of approximately 15 J
for the PDL, and approximately 125 J for the YAG. This is difficult for the laser to achieve and
failure is imminent because the lost power is overheating some component. Good delivery
system transmission is imperative for proper, long-term resonator performance.
Summary of Contents for Cynergy
Page 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Page 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Page 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Page 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Page 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Page 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Page 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Page 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 115: ...Tabbed Page Final Test Calibration ...
Page 116: ......
Page 118: ......
Page 120: ......
Page 148: ......
Page 149: ...Tabbed Page Drawing Sets ...
Page 150: ......
Page 172: ......
Page 173: ...Tabbed Page Schematics ...
Page 174: ......
Page 176: ...This page intentionally left blank ...
Page 178: ...This page intentionally left blank ...
Page 180: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 184: ...This page intentionally left blank ...
Page 186: ...This page intentionally left blank ...
Page 188: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...