Resonator Optics Inspection/Replacement Procedure, PDL
1.
Unplug the system.
2.
Remove the fiber block.
3.
Remove and lay the display assembly module down on the main chassis handle.
4.
Remove the internal monitor assembly.
5.
Inspect the maximum reflector and output coupler carefully for burns or other
degradation. Use a flashlight and compressed air to aid in the inspection. If any damage is
noted, remove the optic with the mount and clean it. First, loosen a compression fitting on
the head for the dye solvent to drain the solvent away—the lines should appear clear
before an optic is replaced.
6.
Remove the four, 4-40 screws that hold the optic to the end plate.
7.
Clean the optic carefully using methanol and a lens tissue or cotton swab. If the optic
cannot be cleaned, replace it.
8.
Install the optic with the mount.
9.
Normally, if the damaged optic was aligned, its replacement should have sufficient
alignment to lase. If the alignment process is being done from scratch, tighten the four
screws that attach the holder to the end plate, and then tighten the three screws that align
the optic using the compliance of the O-ring. Fully compress the O-ring, and then back
off one turn for adjustment on each of the three screws.
Pump Chamber Replacement Procedure
CAUTION: Before removing the pump chamber, always remove the lasers AC line cord
from the wall receptacle. Use a digital voltmeter to make certain that there is no voltage
present on the capacitor bank.
1.
Access the laser resonators.
2.
Remove the rear bridge and disconnect the fan.
3.
Remove the fiber block.
4.
Drain the water.
5.
Disconnect the coolant lines from the pump chamber being serviced. Disconnect the dye
lines for the PDL head. The dye solvent does not have to be drained at this time.
6.
Disconnect the lamp leads from the lamp(s) to be replaced by loosening the 2, 4-40
socket heads on each of the flashlamp collars.
7.
Disconnect the red wire from the trigger transformer that is attached to the pump
chamber.
8.
Remove the rear 10-32 nut that holds the end plate of the resonator to the rubber mount.
9.
Remove the four 4-40 socket head screws that secure the pump chamber to the head
mounting plate. Carefully remove the pump chamber from the laser.
Summary of Contents for Cynergy
Page 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Page 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Page 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Page 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Page 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Page 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Page 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Page 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 115: ...Tabbed Page Final Test Calibration ...
Page 116: ......
Page 118: ......
Page 120: ......
Page 148: ......
Page 149: ...Tabbed Page Drawing Sets ...
Page 150: ......
Page 172: ......
Page 173: ...Tabbed Page Schematics ...
Page 174: ......
Page 176: ...This page intentionally left blank ...
Page 178: ...This page intentionally left blank ...
Page 180: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 184: ...This page intentionally left blank ...
Page 186: ...This page intentionally left blank ...
Page 188: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...