34 of 112
850-1265-000, Rev. 4
Cynosure, Inc.
Rear Control Panel Module
The rear control module,
Figure 11,
consists of the key switch, handpiece filter PCB, bellows
trigger switch, and mounting hardware. The key switch turns the laser on and applies power to
the frame module and front control panel module. The handpiece filter board interfaces with the
laser control board and provides information as to which handpiece is connected to the delivery
fiber. The finger switch signal is also transmitted through this board. The remote interlock, used
to connect to a door switch as specified by CDRH regulations, is resident on this assembly. For
most service operations, the rear control panel stays connected to the chassis. A second piece that
frames the SMA fiber connection is removed to service the laser heads. On this removable piece,
a DC cooling fan pulls hot air out of the chassis, through the rear panel to help maintain the
internal temperature.
Figure 11–Rear Control Panel Module
Summary of Contents for Cynergy
Page 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Page 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Page 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Page 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Page 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Page 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Page 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Page 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 115: ...Tabbed Page Final Test Calibration ...
Page 116: ......
Page 118: ......
Page 120: ......
Page 148: ......
Page 149: ...Tabbed Page Drawing Sets ...
Page 150: ......
Page 172: ......
Page 173: ...Tabbed Page Schematics ...
Page 174: ......
Page 176: ...This page intentionally left blank ...
Page 178: ...This page intentionally left blank ...
Page 180: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 184: ...This page intentionally left blank ...
Page 186: ...This page intentionally left blank ...
Page 188: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...