![Cynosure Cynergy Technical Manual Download Page 15](http://html.mh-extra.com/html/cynosure/cynergy/cynergy_technical-manual_2704655015.webp)
Cynergy Technical Guide
850-1265-000, Rev. 4
13 of 112
Section 2
Equipment Safety
Introduction
As with any electrical equipment, there are potential hazards involved with the operation and
servicing of the laser system. This section of the technical guide identifies these potential hazards
and suggests precautions to avoid them.
Potential Hazards
Optical Hazard
The PDL side of the Cynergy laser generates laser light at a wavelength of 585 nm ± 2% with a
maximum energy of approximately 8 joules delivered from the handpiece. The YAG side
generates laser light at a wavelength of 1064 nm with a maximum energy of approximately 63
joules delivered from the handpiece. Greater energies can be generated from the laser head
especially during service operations. At these wavelengths and energy levels, serious and
permanent damage to the eyes can occur when there is direct or even indirect optical exposure.
WARNING: Do not look directly at laser resonator as the flashlamps fire or severe and
permanent eye damage may occur. These lasers produce laser light each time the
flashlamps fire. Make certain to wear the correct laser eyewear for the wavelength you are
servicing.
Please adhere to the following precautions to avoid optical damage during the operation or
servicing of the laser:
♦
Ensure that everyone present during service procedures wears the appropriate protective
eyewear recommended by Cynosure.
♦
Never look directly into the laser light, even while wearing protective eyewear.
♦
Mark treatment rooms clearly to avoid unexpected entry during treatment or servicing.
♦
Limit entry to the treatment or servicing room to trained, necessary personnel only.
♦
Cover windows and other openings in the treatment room to avoid the inadvertent escape
of laser light.
♦
Cover reflective objects, such as jewelry or mirrors, which could reflect the laser beam to
an area other than the intended treatment area.
♦
Put the laser into the standby mode when the laser is not in use.
♦
Ensure that everyone present during service procedures can shut down the laser in an
emergency.
Summary of Contents for Cynergy
Page 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Page 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Page 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Page 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Page 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Page 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Page 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Page 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 115: ...Tabbed Page Final Test Calibration ...
Page 116: ......
Page 118: ......
Page 120: ......
Page 148: ......
Page 149: ...Tabbed Page Drawing Sets ...
Page 150: ......
Page 172: ......
Page 173: ...Tabbed Page Schematics ...
Page 174: ......
Page 176: ...This page intentionally left blank ...
Page 178: ...This page intentionally left blank ...
Page 180: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 184: ...This page intentionally left blank ...
Page 186: ...This page intentionally left blank ...
Page 188: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...