Water Pump Module
The water pump module,
Figure 20
, consists of the water reservoir/expansion tank, float/level
switch, flow switch, manifold, and the water pump. The water pump receives 230 volts from the
AC distribution PCB. This voltage turns on the water pump and will engage the flow switch if
the system has enough water in the system. The flow switch is rated for 1 gal/min (gal/min),
however, under normal circumstances the water flow should be greater than 2 gal/min. Inside the
expansion tank there is a float or level switch. This switch will indicate a low water level
condition. The operator will need to fill system up until the “low water condition” is reset. The
manifold on the water pump directs the water into the main system and a restricted path to the
DI Cartridge.
Figure 20–Water Pump Module
Summary of Contents for Cynergy
Page 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Page 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Page 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Page 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Page 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Page 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Page 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Page 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Page 115: ...Tabbed Page Final Test Calibration ...
Page 116: ......
Page 118: ......
Page 120: ......
Page 148: ......
Page 149: ...Tabbed Page Drawing Sets ...
Page 150: ......
Page 172: ......
Page 173: ...Tabbed Page Schematics ...
Page 174: ......
Page 176: ...This page intentionally left blank ...
Page 178: ...This page intentionally left blank ...
Page 180: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 184: ...This page intentionally left blank ...
Page 186: ...This page intentionally left blank ...
Page 188: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...