Chapter 3. Using Your ATS
3.11
The Set Target Criteria dialog is displayed, as shown in
.
Figure 3.7: Set Target Criteria Dialog
2. To specify one tolerance for both X and Y measurements, check the box for
x and y the
same
. To specify the tolerance separately for each, uncheck the box. When you check the
box, the dialog changes to display only one set of tolerance criteria for both X and Y.
3. Fill in the tolerances you want to specify. Tolerances must be entered as positive numbers,
must be in millimeters, and must be 0.01 mm to 0.5 mm for X and Y, and 0.1 mm
to1.0 mm for Z.
4. If you don’t want to set a target value for the Z measurement, check the box
ignore z
target value
. Otherwise, uncheck the box and fill in your desired value for
z target value
.
The value must be within two millimeters of the value specified in the target configuration
file (the range, Z ± 1 mm, is shown at the bottom of the dialog).
5. Click
OK
.