The image below shows what the program will look like as a result of the changes for
the FOURTH input folder (above), just verify the white fields are a match as the gray
ones are calculated as part of the calibration:
Click Start.
The values in the gray boxes will be recalculated and again the “Ready” will change
to “Loading data” then to “Analyzing” and then back to “Ready”. Once all of the gray
boxes are filled in again and “Ready” is displayed again in the lower left hand corner
of the box, this input file is completed.
Close the CBGenNormFile application by clicking on the red “X” in the upper right
corner.
Start the Recon Manager Program.
The second, and final, portion of the Panel Air Normalization Calibration is now
complete.
FOURTH Input Folder