14
0473
90001
Rev
07.26.13
Patient Procedure options available for scanning:
Patient Parameters
Procedure
Exposure Factors
Procedure Type
Small Size:
Weight: 50 to 100 lbs (23-45 kg),
Height: G
roin area at least 22” (56
cm) above the floor.
100 kVp, 5mA
Small Patient:
Medium Field
(100kVp)
Small Size:
Weight: 50 to 100 lbs (23-45kg),
Height: G
roin area at least 22” (56
cm) above the floor.
100 kVp, 5mA
Small Patient:
Large Field
(100kVp)
Weight: 101 to 400 lbs (46-181 kg),
Height: G
roin area at least 22” (56
cm) above the floor.
120 kVp, 5mA
Large Field
(120kVp)
Weight: 101 to 400 lbs (46-181 kg),
Height: G
roin area at least 22” (56
cm) above the floor.
120 kVp, 5mA
Medium Field
(120kVp)
Weight: 101 to 400 lbs (46-181 kg),
Height: G
roin area at least 22” (56
cm) above the floor.
120 kVp, 5mA
Medium Field,
Sharp Filter
(120kVp)