7. Mount the Varian Receptor Panel:
a. There are 3 cables to be attached.
CAT6 cable (cable tie this cable to the black clip to maintain security).
Power cable
Ground
cable
b. The Ground nut should be removed from its location, then attach the ground cable
and re-attach the Ground nut.
c. Attach the Panel with 4 Plate screws.
The Load position of the panel is in the Gantry “B” position, which is centered.
Begin by securing the first 2 screws (vertically) and then manually move the
panel to the Gantry Home position, which is rotated to the right, with your hands
and secure the other 2 screws (vertically).