Emergency Removal of a Patient:
The system has undergone extensive testing of the mechanically, electrically and software performance,
but if an unexpected occurrence is observed and/or the software locks up during a scan or
an
emergency arises where it becomes necessary to interrupt a scan and/or remove a patient from
the system before a scan is completed, please follow these steps:
1. Press the EMERGENCY STOP button. This will halt the X-ray as well as the motors
to the machine functions and all power to the machine. The message below will display
on screen and will terminate the “CB Scanning Device” Acquisition software:
2. Open the Gate doors manually with your hands by using the slots at the top of each
door to grasp and slide to an open position.
3. Carefully assist the patient to step out of the scan platform area.
4. Reset the machine: Close the PedCAT Acquisition, “CB Scanning Device” software
(if not already). Release the E-stop by turning the knob to the right until it pops up.
Then turn the machine power back ON at the Main Circuit Breaker. Re-launch the
PedCAT Acquisition “CB Scanning Device” software. Now the system can be operated
again as expected.