1. Find Center of Rotation
Remove all objects from the Field of View in the scanner.
Remove the Patient Platform from the scanner.
Place the Centering Tool
into the center of the metal disc from where
the patient platform was removed. There are 2 holes on the metal disc that will fit the
two metal prongs at the bottom of the centering tool. The Centering Tool will fit in
one of two ways, either way is acceptable for this procedure.
Holes in metal disc in the patient
platform area
(with patient platform removed):
Center tool properly placed:
Double click on the “AlignPanel – Shortcut” on the desktop
to start the
Panel Calibrations.