Power Requirements:
The Scanner requires a Dedicated Line and/or Filtered Line. A Surge Protector is
recommended. The Scanner is suitable for continuous connection to a power supply in stand-by
mode.
Line Voltage
: 100VAC, 115VAC, 200VAC or 230VAC (Factory Set)
Line Voltage Regulation requirement:
± 10%
Line Current:
10 Amps (100V), 10 Amps (115V), 5 Amps (200V) or 5 Amps (230V)
Line Frequency:
50 Hz / 60 Hz
Phase:
Single
Main Circuit Breaker:
10 Amps (100V), 10 Amps (115V), 5 Amps (200V), or 5 Amps
(230V)
Nominal Electrical Input Power to Supply:
Volume Scan = 300W (120kV, 5mA); Scan
Time has no effect on electrical power output.
Apparent Resistance of Supply Mains:
For the purpose of obtaining the apparent resistance of supply mains, resistance is determined
according to the following formula:
R= UO – U1
I1
Where:
U0 is the no-load Mains Voltage
U1 is the Mains Voltage under load.
I1 is the Mains Current under load.
Circuit
Breaker
Assembly
UO
UI
I1
Apparent
Resistance
100VAC
100VAC
98.8VAC
2.5A
0.48ohms
115VAC
115.4VAC
114.2VAC
2.1A
0.57ohms
200VAC
200.4VAC
198.1VAC
1.3A
1.77ohms
230VAC
230.8VAC
228.0VAC
1.2A
2.33ohms
Summary of Contents for pedCAT
Page 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Page 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Page 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Page 97: ...96 0473 90001 Rev 07 26 13 ...
Page 144: ...143 0473 90001 Rev 07 26 13 ...
Page 168: ...167 0473 90001 Rev 07 26 13 ...