105
0473
90001
Rev
07.26.13
Alignment Light Positioning Guides:
There are 3 laser alignment lights to also assist the user in positioning the patient if
needed. Access to turning these lights ON are on the upper corner of the left Door.
When this button is pushed, the 3 lights will illuminate until the button is released.
One is a vertical light in the very center of rotation (center of the circles). The anatomy
that this light falls on will be in the very center of the scan.
The other two lights are horizontal, one at the top of the scan (20 cm height), the other at
the center of the scan (10 cm height).
Patient Instructions for a Scan:
Once the patient is properly positioned in the system, the operator should instruct the patient
to
hold perfectly still for the duration of the scan
. It is very important for the patient to
hold still. If the patient moves during the volume scans, the results may not be optimal.
Full Both feet procedure times will be approx. 68 seconds
Single foot procedure times will be approx. 19 seconds
WARNING
Laser beams can cause optical damage. The operator
should avoid looking directly into the beams. The operator should instruct
the patient to avoid looking directly into the beams. The use of optical
instruments such as eyeglasses with large diopter or mirrors, increase eye
hazard with this product.
Alignment
Lights
button
Summary of Contents for pedCAT
Page 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Page 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Page 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Page 97: ...96 0473 90001 Rev 07 26 13 ...
Page 144: ...143 0473 90001 Rev 07 26 13 ...
Page 168: ...167 0473 90001 Rev 07 26 13 ...