97
CU-SP2 Battery Pack (Rechargeable)
Model
CUA1802RB (3INR/19/65)
Battery Type
10.8VDC, 2.5Ah Li-ion, rechargeable
Capacity
Charging Time with
device off
For fully charged new batteries, at least 60 shocks or 3
hours of operation at 25°C (77°F)
Approximately 3 hours to 100%. Approximately
2.8 hours to 90%.
Charging the battery at temperatures above 45℃ may
degrade battery life. (for standard battery)
Standby Life (After
Inserting the Battery)
If stored and managed in accordance with instructions in
the document:
At least 2 years from the date of installation into the i-PAD
CU-SP2
Temperature Ranges for
Storage and Use
Operating Environment
Temperature: 0°C ~ 40°C (32°F ~ 104°F)
Storage Environment
Temperature: -20°C ~ 60°C (-4°F ~ 140°F)
CU-SP2 Battery Pack (Disposable)
Model
CUSA1103BB
Battery Type
12V DC, 4.2Ah LiMnO
2
, disposable
Capacity
For fully charged new batteries, at least 130 shocks or 5
hours of operation at 25°C (77°F)
Standby Life (After
Inserting the Battery)
If stored and managed in accordance with instructions in the
document:
At least 5 years from the date of installation into the i-PAD
CU-SP2
Temperature Ranges for
Storage and Use
Operating Environment
Temperature: 0°C ~ 43°C (32°F ~ 109°F)
Storage Environment
Temperature: -20°C ~ 60°C (-4°F ~ 140°F)