54
The second tab displays the treatment data saved on the device. For further details on 'DATA
REVIEW' on the second tab, please refer to [Section 5.3: Data Review] in these Instructions for
Use.
The third tab is for exiting Menu Mode.
If pads are attached to the patient while the device is in Menu Mode, the device will
automatically switch from Menu Mode into Operation Mode to enable defibrillation.
5.4.2 Setting the Operation of the Device
The user can set the options below under the 'DEVICE SETUP' tab of Menu Mode.
Device Mode
•
With the security mode activated, you should enter the password to change the Device
Mode.
The default password of the device is Soft Key ‘
1
1
1
1’
. (Soft Keys are labeled
1~3 from left to right.)
•
AED Mode:
The device manually or automatically executes ECG analysis and defibrillation.
•
Manual Mode:
The user can manually set the device's shock energy and administer
defibrillation based on user judgment.