22
Button 2 is activated only when the device is set to Pediatric Mode.
When 'Manual Override' is set to 'OFF', Soft Keys 1 and 3 will be deactivated after attaching
the pads on the patient.
4.3 Mode Change
The operation mode change is accomplished by pressing the i-button while the device is on.
Soft Keys 2 and 3 are activated by pressing the i-button. By pressing the i-button and a soft key
at the same time, operation mode will be changed.
Soft keys for mode changes are as follows:
※
Soft Keys are labeled 1~3 from left to right.
Soft Key Functions in Operation Mode (Pressed the i-button to change the mode)
Button
Indication
Function
Button 2
Adult
This function is activated when the i-button is pressed while the
device is in pediatric mode.
When pressed, mode is changed to adult mode.
Pediatric
This function is activated when the i-button is pressed while the
device is in adult mode.
When pressed, mode is changed to pediatric mode.
Button 3
Manual
Mode
This function is activated when the i-button is pressed while the
device is in AED mode.
When pressed, mode is changed to Manual mode.
AED Mode
This function is activated when the i-button is pressed while the
device is in Manual mode.
When pressed, mode is changed to AED mode.
Button 3 is activated only when manual mode is installed. When the equipment security mode
setting is activated, changing the mode asks you to enter the password. For further details on
the security mode, please refer to [Section 5.4: Device Setup] in the Instructions for Use.
When pediatric pads are connected, the button 2 is not activated.