4.5 Preparation for Defibrillation
①
Turn the device on by pressing the Power Button.
When the power turns ON the following occurs in sequence:
Beeper: The beeper will beep for 1 second.
A self-test will be initiated.
The device will give voice instructions to call emergency medical
services and on the 'Adult / Pediatric Mode'.
The guide on how to use the device will be given through the LCD
Screen and by voice.
Never perform defibrillation in pediatric mode to a patient who is either heavier than 25 kg or
older than 8 years old.
You can change the adult/pediatric mode under Menu Mode after turning on the i-PAD CU-
SP2. However, the defibrillation mode should be changed before placing the pads on the
patient. Once the pads are in place, you cannot change the defibrillation mode anymore.
When the mode is correctly selected, the defibrillation energy is set to an adult value (150 J /
200J) or pediatric value (50 J).
For further details on setting the menu, please refer to [Section 5.4: Device Setup] of these
Instructions for Use.