ECG Accuracy
Category
General Specifications
ECG Acquisition Route
Lead II
Response Frequency
1 Hz ~ 30 Hz
ECG Analysis System
Category
General Specifications
Function
Analyzes whether the rhythms of the patient's impedance and
ECG require a defibrillation
Measured Impedance
Range
25Ω ~ 175Ω
Rhythm Requiring
Defibrillation
Ventricular fibrillation and several ventricular tachycardia
including ventricular flutter
The CU-SP2 uses multiple variables to determine the
shockability of the heartbeat.
Some extremely low amplitudes or low frequency heartbeats
are not interpreted as shockable VF beats. Also, some VT
beats are not interpreted as rhythms requiring defibrillation.
Rhythm Not Requiring
Defibrillation
ECG rhythms excluding those requiring a defibrillation
When a rhythm that does not require a defibrillation is detected,
the device informs the user by voice to perform CPR.
Analysis Protocol
Prepares to administer shock or give voice instructions on CPR
according to the analysis result
Algorithm sensitivity
and specifications that
require defibrillation
Satisfies AAMI DF80