46
③
If the i-Button is pressed for more than 1 second in standby mode, the device will switch to
Administration Mode and give instructions by voice and LCD Screen.
④
The device displays the summary (the total hours of the last device usage and the number of
defibrillation shocks delivered) of the device usage on the LCD Screen.
⑤
The device displays the S/W version on the LCD Screen.
⑥
When the voice guide instructs to transfer the treatment data, press the i-Button to copy the
data onto the SD card.
If there is treatment data in the device’s internal memory:
The device starts to copy the data after informing the user by voice that the treatment data is
being copied onto the SD card.
When copying is completed, the device informs you by voice and automatically shuts down.
If there is no treatment data in the device’s internal memory:
The device informs you by voice that no treatment data exists and automatically shuts down.
If the SD card already has the same treatment data file, the device informs the user that the
same file already exists upon copying the treatment data onto the SD card. Press the Shock
Button to overwrite the existing file or press the i-Button to cancel copying the file.