41
When pressing Soft Key 1, the output energy value displayed on the upper-right corner of the
LCD Screen will be highlighted in white. Also, the Soft Keys will be activated as follows:
•
Soft Key 1: ▲
(INCREASE ENERGY)
•
Soft Key 2:
▼ (DECREASE ENERGY)
•
Soft Key 3:
CONFIRM
At this time, use Soft Keys 1 and 2 to change the energy value and press Soft Key 3 to confirm.
[Output energy values for Adult / Pediatric Mode]
Adult / Pediatric Mode
Output Energy
Adult
2J, 3J, 5J, 7J, 10J, 20J, 30J, 50J, 70J, 100J, 150J, 200J
Pediatric
2J, 3J, 5J, 7J, 10J, 20J, 30J, 50J
4.11.2 Charging the Device and Administering Electric Shock Treatment
In manual mode, the user may, in his/her own discretion, charge the i-PAD CU-SP2 and
administer defibrillation.
After attaching the pads on the patient, press ‘CHARGE’ using Soft Key 3 if the patient’s ECG is
shockable. When pressing Soft Key 3, the device will start charging according to the set energy
level along with a charging sound. The charged energy amount can be checked on the LCD
Screen.