96
Control Devices, Indicators, Voice Instructions
Category
General Specifications
Control Devices
Power Button, i-Button, Shock Button, 3 Soft Keys
Graphic LCD
Displays the operating status of the device and instructions
Indicators
Shock Button: Flashes in orange when the defibrillator is charged
and ready to deliver a shock.
Blue i-Button: Flashes when guiding CPR, transferring treatment
data or setting the CPR mode.
Red i-Button: Flashes when an error occurs.
Speakers
Outputs voice instructions
If the device determines, based on its settings, that the surrounding
environment is noisy and it cannot give accurate voice instructions,
it will automatically increase the volume for the user.
Beeper
Outputs various beeps
Low Battery Check
The check is automatically performed through periodical self-tests as
well as in real-time when the equipment is in use or the power is
turned on.
Low Battery
Indicator
The Graphic LCD on the device indicates low battery along with voice
instructions and a flashing red i-Button.
Voice Instruction
Detailed voice instructions during defibrillation and CPR.
Self-Tests
Auto
Power Self-Test / Real-time Self-Test
Daily / Weekly / Monthly Self-Test
Manual
Battery Pack Self-Test (performed when the user inserts the battery pack)