91
Arming Indicator
Voice instruction (Press the flashing orange button.)
Flashing Shock Button
Beeper
Time from End of
CPR to Administering
Shock
At least 6 seconds from the completion of CPR to the shock delivery
Disarm
The device disarms the electric load under the following situations:
•
When the patient's ECG is changed into a status that does not
require defibrillation.
•
When the Shock Button is not pressed within 15 seconds from the
completion of the charge.
•
When the equipment is turned off by pressing the Power Button for
over 1 second.
•
When the pad is detached from the patient's body or the pads
connector is detached from the device.
•
When the impedance of the patient is out of the range of
defibrillation.
(25Ω ~ 175Ω)
Electric Shock
After charging is completed, the device delivers a defibrillating shock
to the patient when the Shock Button is pressed.
Vector for
Administering Shock
The pads (Lead II) are placed anterior-anterior for the adult.
The pads are placed anterior-posterior for the child.
Patient Isolation
BF Type, defibrillation protected
Sync
Delay between synchronization
pulse and shock delivery
The shock is delivered within 60 milliseconds of the
occurrence of a QRS peak in the patient’s ECG