34
4.9 Monitor Mode
Monitor Mode is used in connection with the CU-EM1 (ECG transmission device). In Monitor
Mode, the i-PAD CU-SP2 uses Bluetooth to receive ECG data from the CU-EM1 and displays it
on the LCD Screen. When using Monitor Mode, the pads cannot be used and defibrillation
cannot be performed. If you think that defibrillation is necessary while using Monitor Mode on
the patient, immediately turn Monitor Mode off and administer defibrillation.
To use the CU-EM1, you must first pair the CU-SP2 and the CU-EM1 in Menu Mode. For further
details on pairing the CU-EM1, please refer to [Section 5.4: Device Setup] in these Instructions
for Use.
The CU-EM1 needs to be paired only once, and will be automatically connected in the future.
However, you will need to do pairing again for a different CU-EM1.
To use Monitor Mode, press Soft Key 1, which reads 'MONITOR MODE', without attaching the
pads on the patient.
When pressed, the device will attempt to connect with the CU-EM1.