The usage authority differs for each device mode.
Manual Mode
: Only medical professionals may use this mode.
AED Mode
•
ANALYZE
: Only licensed emergency medical technicians or medical professionals may use
this mode.
•
CHARGE
: Only medical professionals may use this mode.
•
OFF
: Licensed emergency medical technicians, medical professionals and the general
public may use this mode.
In case where the patient’s ECG signal is in the asystole state, the asynchronous
defibrillation therapy may lead to a failure to restore cardiac pacemaker function so the
cardiac function will not be restored. Therefore, do not deliver electric shock to
patients with asystole.
When administering R-Sync energy, the patient may be administered with defibrillation
energy, recognized as R-waves, if there is interference resulting from external contact
or if the patient is moved while the pads are attached. Avoid moving or touching the
patient while administering R-Sync energy.
Manual Mode is an additional option. If you want to add the manual mode option, please
contact us after referring to [Section A.3: Service Center] in these Instructions for Use.