90
Defibrillator
Category
General Specifications
Operation Type
Dual Mode (semi-automated, manual) External Defibrillator
Output Type
e-cube biphasic (Truncated exponential type)
Output Energy
AED Mode
•
150J±4J, 200J±6J
at 50Ω load for adults
•
50J±2J
at 50Ω load for children
Manual Mode (Optional)
•
2J±1J, 3J±1J, 5J±1J, 7J±1J, 10J±1J, 20J±2J, 30J±2J, 50J±2J,
70J±2J, 100J±4J, 150J±4J, 200J±6J at
50Ω load
Charge Control
Controlled by an automated patient analysis system
Charge Time
Manual Mode
•
Charging time, set to maximum energy output, new
fully charged battery pack
New battery pack
12 Seconds, typical
•
Charging time, set to maximum energy output, 15
th
shock discharge from a new fully charged battery
pack
New battery pack: 15
th
shock
discharge 14 Seconds,
typical
•
Time from initially switching power ON to ready for
discharge, set to maximum energy output, 15
th
shock
discharge from a new fully charged battery pack
New battery pack: 15
th
shock
discharge 24 Seconds,
typical
AED Mode
•
Time from initiation of rhythm analysis to readiness for
discharge, set to maximum energy output, new fully
charged battery pack
New battery pack
12 Seconds, typical
•
Time from initiation of rhythm analysis to readiness for
discharge, set to maximum energy output, 15
th
shock
discharge from a new fully charged battery pack
New battery pack: 15
th
shock
discharge 14 Seconds,
typical
•
Time from initially switching power ON to ready for
discharge, set to maximum energy output, 15
th
shock
discharge from a new fully charged battery pack
New battery pack: 15
th
shock
discharge 25 Seconds,
typical