89
D. Device Specifications
Model Name: CU-SP2
Product Exterior
Category
General Specifications
Dimensions
260
㎜
x 256
㎜
x 69.5
㎜
(Width x Length x Height)
Weight
2.4
㎏
(Including the battery pack and pads)
Environmental Conditions
Category
General Specifications
Operating Environment
(The device can be used immediately in case of an emergency.)
Temperature
: 0°C ~ 40°C (32°F ~ 104°F)
Humidity
: 5% ~ 95% (a location with no condensation)
Storage Environment
(The device has pads and a battery and is ready to be used for an
emergency.)
Temperature
: 0°C ~ 43°C (32°F ~ 109°F)
Humidity
: 5% ~ 95% (a location with no condensation)
Transportation Environment
(The device does not have pads and a battery and is
separately stored or transported over a long period of time.)
Temperature
: -20°C ~ 60°C (-4°F ~ 140°F)
Humidity
: 5% ~ 95% (a location with no condensation)
Altitude
0 to 5,000 m (operational and storage)
Drop
Withstands 1.2-meter drop to any edge, corner, or surface
Vibration
Operating: Meets MIL-STD-810G Fig.514.6E-1, random
Standby: Meets MIL-STD-810G Fig.514.6E-2, swept
sine(helicopter)
Sealing
IEC 60529: IP55
ESD
Meets IEC 61000-4-2:2001
EMI (Radiated)
Meets IEC 60601-1-2 limits, method EN 55011:2007 +A2:2007,
Group 1, Class B
EMI (Immunity)
Meets IEC 60601-1-2 limits, method EN 61000-4-3:2006 +A1:2008
Level 3 (10V/m 80MHz to 2500MHz)