![background image](http://html1.mh-extra.com/html/covidien/valleylab-ft-series/valleylab-ft-series_user-manual_2673869063.webp)
Monopolar
Valleylab FT10 Energy Platform User’s Guide
4-5
Monopolar Function Overview
Monopolar Function Overview
Monopolar Power Output Modes
The monopolar energy channels can be configured to these modes: PURE CUT, BLEND,
VALLEYLAB, SOFT, FULGURATE, and SPRAY.
Foot Pedal
Monopolar foot pedals connect to the rear panel and correspond to either the Monopolar
1 or Monopolar 2 receptacles on the front panel.
Standard foot-pedal devices attached to the Monopolar 1 Universal Foot Pedal Port
receptacle on the front panel can be controlled with a Covidien foot pedal attached to the
Monopolar 1 foot-pedal receptacle on the back panel.
Warning
Electric Shock Hazard
•
Do not connect wet instruments to the system.
•
Ensure that all instruments and adapters are correctly connected and that no metal is exposed
at any connection point.
Connect instruments to the proper receptacle. Improper connection may result in inadvertent
instrument activation or other potentially hazardous conditions. Follow the instructions provided
with electrosurgical instruments for proper connection and use.
Each instrument receptacle on this energy platform is designed to accept only one instrument at a
time. Follow the instructions provided with electrosurgical instruments for proper connection and
use.
Precaution
Read the instructions, warnings, and precautions provided with electrosurgical instruments before
use. Specific instructions are not included in this manual.
Inspect instruments and cords—especially for laparoscopic/endoscopic instruments—for breaks,
cracks, nicks, and other damage before every use. If damaged, do not use. Damaged instruments
or cords may result in injury or electrical shock to the patient or surgical team.
Important
Avoid switching rapidly between CUT and COAG with the monopolar handpiece controls as this
may cause the generator to temporarily ignore the handpiece control. Release the handpiece
control momentarily and then press it again to rectify this situation.