Technical Specifications
Valleylab FT10 Energy Platform User’s Guide
10-17
Standards and IEC Classifications
Frequency ranges and text condition on transmitter bands
Bands (MHz)
Test Frequencies
(MHz)
Modulation
Claimed Compliance
Level V/m
380 – 390
385
Pulse
a
– 18 Hz
27
430 – 470
450
FM ± 5 kHz Deviation
or Pulse
a
– 18 Hz
28
704 – 787
710, 745, 780
Pulse
a
– 217 Hz
9
800 – 960
810, 870, 930
Pulse
a
– 18 Hz
28
1700 – 1990
1720, 1845, 1970
Pulse
a
– 217 Hz
28
2400 – 2570
2450
Pulse
a
– 217 Hz
28
5100 – 5800
5240, 5500, 5785
Pulse
a
– 217 Hz
9
a.
Pulse Modulation is defined as a square wave input with a 50% duty cycle at the listed frequency.
NOTE 1
The minimum dwell time during test is 7 s.
NOTE 2
A minimum separation distance of 0.3 meters should be maintained between any device
transmitting in this band and the VLFT10GEN. This includes devices such as mobile phones, PDAs,
Wireless LANs, RFID and Bluetooth™.
Recommended separation distances between portable and mobile RF communication
equipment and the Valleylab FT10 FT Series Energy Platform