Tr
oubleshooting
Valleylab FT10 Energy Platform User’s Guide
8-13
Correcting Malfunctions
E310
TOUCHSCREEN
ERROR
Unintentional
Screen Input
Medium Alarm
Condition clears
E312
INTERNAL CHECK
Absolute
Integrated Error
Medium Alarm
Reactivation, User
acknowledge, Enter
Service
E313
INTERNAL CHECK
Voltage Sensor
Compare Error
Medium Alarm
Reactivation, User
acknowledge, Enter
Service
E314
INTERNAL CHECK
Current Sensor
Compare Error
Medium Alarm
Reactivation, User
acknowledge, Enter
Service
E315
INTERNAL CHECK
Power Sensor
Compare Error
Medium Alarm
Reactivation, User
acknowledge, Enter
Service
E316
INTERNAL CHECK
Power
Compare Error
Medium Alarm
Reactivation, User
acknowledge, Enter
Service
E317
INTERNAL CHECK
PVI Curve
Compare Error
Medium Alarm
Reactivation, User
acknowledge, Enter
Service
E318
INTERNAL CHECK
Voltage Sensor
Clip Error
Medium Alarm
Reactivation, User
acknowledge, Enter
Service
E319
INTERNAL CHECK
Current Sensor
Clip Error
Medium Alarm
Reactivation, User
acknowledge, Enter
Service
E320
CHECK
INSTRUMENT
N/A
Medium Alarm
Switch release,
Remove instrument
E321
CHECK FOOT
PEDAL
N/A
Medium Alarm
Pedal release
E322
CHECK
INSTRUMENT
N/A
Medium Alarm
Switch release,
Remove instrument
E323
CHECK FOOT
PEDAL
N/A
Medium Alarm
Pedal release
E324
CHECK
INSTRUMENT
N/A
Medium Alarm
Switch release,
Remove instrument
E325
CHECK FOOT
PEDAL
N/A
Medium Alarm
Pedal release
Error
Code
Error Title
Description
Alarm Type
Clearing Condition