![Covidien Valleylab FT Series User Manual Download Page 120](http://html1.mh-extra.com/html/covidien/valleylab-ft-series/valleylab-ft-series_user-manual_2673869120.webp)
8-16
Valleylab FT10 Energy Platform User’s Guide
Correcting Malfunctions
E401
AUTHORIZATION
ISSUE
Invalid Pairing
Code
Low Alarm
User acknowledge,
Enter Service
E402
INCOMPLETE SEAL
CYCLE
N/A
LigaSure Alarm
Reactivation, Remove
instrument, User
acknowledge, Enter
Service, Install
instrument
E403
INCOMPLETE SEAL
CYCLE
N/A
LigaSure Alarm
Reactivation, Remove
instrument, User
acknowledge, Enter
Service
E404
INCOMPLETE SEAL
CYCLE
N/A
LigaSure Alarm
Reactivation, Remove
instrument, User
acknowledge, Enter
Service
E405
INCOMPLETE SEAL
CYCLE
N/A
LigaSure Alarm
Reactivation, Remove
instrument, User
acknowledge, Enter
Service
E407
INSTRUMENT NOT
READY
N/A
Medium Alarm
Other plug on
instrument inserted
into LigaSure port,
Remove instrument
E408
INSTRUMENT NOT
READY
N/A
Medium Alarm
Other plug on
instrument inserted
into Mono 2 port,
Remove instrument
E411
INVALID
ACTIVATION
ATTEMPT
N/A
Medium Alarm
Insert other half of
plug into LigaSure
port, Remove
instrument
E412
INVALID
ACTIVATION
ATTEMPT
N/A
Medium Alarm
Insert other half of
plug into Mono 2 port,
Remove instrument
E413
INVALID
COMBINATION
N/A
Medium Alarm
Remove instrument.
Insert other half of
plug into LigaSure port
E414
INVALID
COMBINATION
N/A
Medium Alarm
Remove instrument.
Insert other half of
plug into Mono 2 port
Error
Code
Error Title
Description
Alarm Type
Clearing Condition