Technical Specifications
Valleylab FT10 Energy Platform User’s Guide
10-23
Energy Output Characteristics
Energy Output Characteristics
The VLFT10GEN automatically senses the tissue resistance and adjusts the output voltage
to maintain a consistent effect across different tissue density. The adjustment is based on
the selected mode or effect, power setting, and the level of tissue resistance. The
maximum output voltage is controlled to reduce capacitive coupling and video
interference and to minimize undesired sparking. The following table outlines the
maximum power, peak voltages, and maximum output current for a given modality.
Notice
Alarm characteristics cannot be changed.
Alarm response time is less than one second.
Rated
Load (Ω)
Rated
Output
Power
(W)
Peak
Voltage
Current
RMS
Max
Typical
Crest
Factor
Duty
Cycle
Monopolar CUT
PURE
BLEND
VALLEYLAB
300
300
300
300
200
200
1287
2178
2783
1.25
1
1
1.42
2.5
3.8
100%
50%
25%
Monopolar COAG
SOFT
FULGURATE
SPRAY
100
500
500
120
120
120
264
3449
3933
1.55
1
1
1.42
5.7
6.5
100%
6.25%
4.76%
Bipolar
LOW (1–15 W)
MEDIUM (16–40 W)
HIGH (45–95 W)
100
100
100
15
40
95
133
214
462
1
2
2
1.42
1.42
1.42
100%
100%
100%
LigaSure
LIGASURE
BIPOLAR
RESECTION
20
350
244
5.5
1.42
100%
•
CUT
•
COAG
500
100
200
175
742
318
2.4
3.2
1.42
1.42
100%
100%