![background image](http://html1.mh-extra.com/html/covidien/valleylab-ft-series/valleylab-ft-series_user-manual_2673869148.webp)
10-18
Valleylab FT10 Energy Platform User’s Guide
Standards and IEC Classifications
Cables Used for EMC Compliance Testing
The following cables, lengths, and instruments were used to determine compliance levels.
Model No.
Description
Shielded
Length
Location
Classification
Tests
Included
207002060
A/C cable
No
15 ft.
(4.6 m)
Rear
panel
AC
Emissions/
Immunity
1017178
Equipotential
cable
No
15 ft.
(4.6 m)
Rear
panel
EQ
Emissions/
Immunity
E2450HDB
2-button
pencil
No
15 ft.
(4.6 m)
Front
panel
Mono 1
Patient
Emissions/
Immunity
FT3000DB
Force
TriVerse
pencil
No
15 ft.
(4.6 m)
Front
panel
Mono 2
Patient
Emissions/
Immunity
E7507DB Adult
REM
pad
No
15 ft.
(4.6 m)
Front
panel
REM
Patient
Emissions/
Immunity
E0020V/
E4073CT
Foot
switching
Bipolar cord/
bayonet
forceps
No
15 ft.
(4.6 m)
Front
panel
Bipolar
Patient
Emissions/
Immunity
LF1644
LigaSure
No 10
ft.
(3 m)
Front
panel
LigaSure
Patient Immunity
FT0021S/
FT27050SL
Cord Storz/
Storz
Resectoscope
No
15 ft.
(4.6 m)/
NA
Front
panel
Adv.
Energy
Patient
Emissions
E6008B
Monopolar
footswitch
No
15 ft.
(4.6 m)
Rear
panel
Mono 1
SIP/SOP
Emissions/
Immunity
E6019
Dome Bipolar
footswitch
No
15 ft.
(4.6 m)
Rear
panel
Bipolar
SIP/SOP
Emissions/
Immunity
LS0300
LigaSure
footswitch
No
15 ft.
(4.6 m)
Rear
panel
LigaSure
SIP/SOP
Emissions/
Immunity
Continued