Tr
oubleshooting
Valleylab FT10 Energy Platform User’s Guide
8-17
Correcting Malfunctions
E415
FEATURE NOT
ENABLED
N/A
Medium Alarm
Remove instrument,
Enter Service, Dismiss
message (select Okay
or Cancel)
E416
UNKNOWN
INSTRUMENT
N/A
Medium Alarm
Remove instrument
E417
INVALID
INSTRUMENT
N/A
Medium Alarm
Remove instrument
E419
Incomplete Setup
N/A
Low Alarm
Remove instrument.
Enter Service, EM
clearing message, RF
activation
E420
Usage Limit
N/A
Medium Alarm
Replace instrument
E421
Usage Limit
N/A
Medium Alarm
User acknowledge, RF
activation
E422
Incomplete Setup
N/A
Low Alarm
Remove instrument,
Enter Service, EM
clearing message, RF
activation
E424
Activation Time
Exceeded
N/A
Medium Alert
Remove instrument.
Enter Service. EM
clearing message
(open circuit or open
circuit timeout), RF
activation
E425
Contact time
Exceeded
N/A
Medium Alert
Remove instrument,
Enter Service, EM
clearing message
(open circuit), RF
activation
Error
Code
Error Title
Description
Alarm Type
Clearing Condition