10-20
Valleylab FT10 Energy Platform User’s Guide
Standards and IEC Classifications
Auto Bipolar
The VLFT10GEN is equipped with an Auto Bipolar feature that allows for automatic
activation of bipolar energy.
Warning
Use of different Covidien cord models or cords from other manufacturers may not achieve proper
electrical output for this device, thereby failing to produce the desired clinical effect. For example,
Auto Bipolar activation/deactivation settings may not work properly using cords other than those
specified by Covidien.
In the Auto Bipolar setting, activation may occur with contact of any material. When not in use,
place electrosurgical instruments in a safety holster or safely away from patients, the surgical team,
and flammable materials
Auto Bipolar
Specifications
Duration
Interrogation Frequency
434 kHz
Interrogation Current
< 10
μA
RMS
averaged over 1 second
Interrogation Voltage
< 12 V
RMS
Activation Impedance
2200 Ω ± 20%
Deactivation Impedance
> 4000 Ω ± 20%
Keying Delay
User selectable in 0.5 second increments from 0 sec. to 2.5 sec.
Minimum Power
1 W