8-14
Valleylab FT10 Energy Platform User’s Guide
Correcting Malfunctions
E326
CHECK
INSTRUMENT
N/A
Medium Alarm
Switch release,
Remove instrument
E327
CHECK FOOT
PEDAL
N/a
Medium Alarm
Pedal release
E328
UNINTENTIONAL
INPUT
Panel Button
Error
Medium Alarm
Button release
E329
CHECK FOOT
PEDAL
Invalid Foot
Pedal Input
Medium Alarm
Condition clears
E330
PROLONGED
ACTIVATION
Prolonged
Activation
Medium Alarm
User acknowledge,
Enter Service,
Reactivation
E331
INVALID
INSTRUMENT
Invalid
Instrument
Medium Alarm
Instrument removal
E332
SYSTEM ISSUE
Unit Overtemp
Detected
Medium Alarm
User acknowledge,
Enter Service
E333
SYSTEM ISSUE
Cooling Fan
Issue
Medium Alarm
User acknowledge,
Enter Service
E334
REM CHECK ISSUE
REM
Automated
Check Error
Medium Alarm
Successful completion
of REM check, User
acknowledge, Enter
Service
E335
SYSTEM ISSUE
Driver Comm
Issue
Medium Alarm
User acknowledge,
Enter Service
E336
INTERNAL CHECK
ADC Sync Issue
Medium Alarm
Start RF activation,
User acknowledge,
Enter Service
E337
INTERNAL CHECK
Real Time Clock
Issue
Medium Alarm
User acknowledge,
Enter Service
E339
INTERNAL CHECK
RF Control Limit
Check Issue
Medium Alarm
Start RF activation,
User acknowledge,
Enter Service
E340
CONFIGURATION
MISMATCH
Invalid Country
Code
Medium Alarm
Connect to VLEX
E341
INTERNAL CHECK
Panel Comm
Issue
Medium Alarm
Condition clears, User
acknowledge, Enter
Service
Error
Code
Error Title
Description
Alarm Type
Clearing Condition