8-10
Valleylab FT10 Energy Platform User’s Guide
Correcting Malfunctions
E208
SYSTEM ERROR
Stack Overflow
High Alarm
Power on reboot,
successful POST
E209
SYSTEM ERROR
DMA Failure
High Alarm
Power on reboot,
successful POST
E210
SYSTEM ERROR
Thread Limit
Error
High Alarm
Power on reboot,
successful POST
E211
SYSTEM ERROR
Internal
Messaging
Error
High Alarm
Power on reboot,
successful POST
E212
SYSTEM ERROR
PCBA Failure
High Alarm
Power on reboot,
successful POST
E213
SYSTEM ERROR
SR PCBA Failure
High Alarm
Power on reboot,
successful POST
E214
SYSTEM ERROR
RF PCBA Failure
High Alarm
Power on reboot,
successful POST
E215
SYSTEM ERROR
Software State
Error
High Alarm
Power on reboot,
successful POST
E216
SYSTEM ERROR
Low Voltage
Rail Failure
High Alarm
Power on reboot,
successful POST
E217
SYSTEM ERROR
High Voltage
Rail Failure
High Alarm
Power on reboot,
successful POST
E218
SYSTEM ERROR
Display Driver
Error
High Alarm
Power on reboot,
successful POST
E219
SYSTEM ERROR
Audio Circuit
Failure
High Alarm
Power on reboot,
successful POST
E220
SYSTEM ERROR
Multi-tasking
Error
High Alarm
Power on reboot,
successful POST
E221
SYSTEM ERROR
Processor
Timebase Error
High Alarm
Power on reboot,
successful POST
E222
SYSTEM ERROR
Processor
Timebase Error
High Alarm
Power on reboot,
successful POST
E223
SYSTEM ERROR
FPGA Timebase
Error
High Alarm
Power on reboot,
successful POST
E224
SYSTEM ERROR
System
Timebase Error
High Alarm
Power on reboot,
successful POST
Error
Code
Error Title
Description
Alarm Type
Clearing Condition