K-FTH-1012 Instructions for Use
English
1-14
1
Guidance and manufacturer’s declaration – electromagnetic immunity
(continued)
Conducted RF
IEC 61000-4-6
3Vrms
0.15 to 80MHz
6Vrms in ISM bands
Between 0.15MHz
and 80MHz
c
80% AM at 1kHz
3Vrms
0.15 to 80MHz
6Vrms in ISM bands
Between 0.15MHz
and 80MHz
c
80% AM at 1kHz
Portable and mobile RF communications
equipment should be used no closer to
any part of the Test Tube Heater,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommend separation distance
d = 0.6 √P
Recommend separation distance
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
should be
less than the compliance level in each
frequency range
b
. Interference may
occur in the vicinity of equipment
marked with the following symbol:
Radiated RF
IEC 61000-4-3
3V/m
80MHz to 2.5GHz
80% AM at 1kHz
3V/m
80MHz to 2.5GHz
80% AM at 1kHz
Proximity fields from RF wireless communications equipment IEC 61000-4-3.
Note 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the Test Tube Heater is used exceeds the applicable RF compliance level above,
the Test Tube Heater should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the Test Tube
Heater.
b
Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.
c
The ISM (industrial. scientific and medical) bands between 0.15MHz and 80MHz are 6.765MHz to
6.795MHz; 13.553MHz to 13.567MHz; 26.957MHz to 27.283MHz; and 40.66MHz to 40.70MHz. The
amateur radio bands between 0.15MHz and 80MHz are 1.8MHz to 2.0MHz; 3.5MHz to 4.0MHz; 5.3MHz
to 5.4MHz; 7.0MHz to 7.3MHz; 10.1MHz to 10.15MHz; 14MHz to 14.2MHz, 18.07MHz to 18.17MHz;
21.0MHz to 21.4MHz; 24.89MHz to 24.99MHz; 28.0MHz to 29.7MHz; and 50.0 MHz to 54.0MHz.
Summary of Contents for K-FTH-1012
Page 113: ...K FTH 1012 7 3 K FTH 1012 CE 2012 19 K APS 300 CE 2012 19 IP64 EC REP REF SN 7...
Page 114: ...K FTH 1012 7 4 1 7...
Page 115: ...K FTH 1012 7 5 1 1 36 9 C K FTH 1012 Cook 7...
Page 117: ...K FTH 1012 7 7 3 3 100 240 VAC 50 60 Hz 3 4 EMC IEC 60601 1 2 2014 EMC EMI Cook 7 7...
Page 118: ...K FTH 1012 7 8 4 1 1 K FTH 1012 CP 2 3 4 K FTH 1012 5 4 2 1 2 3 4 5 4 7...
Page 120: ...K FTH 1012 7 10 5 4 5 5 6 6 1 70 6 2 4 3 1 2 3 6 5 mL 4 5 6 7...
Page 121: ...K FTH 1012 7 11 1 2 3 6 5 mL 1 2 3 1 0 C 36 9 C 6 3 SMM30004 SMM30004 6 4 7...
Page 204: ...K FTH 1012 12 4 1 12...
Page 205: ...K FTH 1012 12 5 1 1 36 9 C K FTH 1012 Cook 12...
Page 207: ...K FTH 1012 12 7 3 3 100 240 50 60 3 4 EMC IEC 60601 1 2 2014 EMC EMI Cook 7 12...
Page 208: ...K FTH 1012 12 8 4 1 1 K FTH 1012 CP 2 3 4 K FTH 1012 5 4 2 1 2 3 4 5 4 12...
Page 210: ...K FTH 1012 12 10 5 4 5 5 6 6 1 70 6 2 4 3 1 2 3 6 5 4 5 6 12...
Page 211: ...K FTH 1012 12 11 1 2 3 6 5 1 2 3 1 0 C 36 9 C 6 3 SMM30004 SMM30004 6 4 12...
Page 291: ...Notes...
Page 292: ...Notes...
Page 293: ...Notes...