K-FTH-1012 Instructions for Use
English
1-13
1
Guidance and manufacturer’s declaration – electromagnetic immunity
The Test Tube Heater is intended for use in the electromagnetic environment specified below. The
customer or the end user of the Test Tube Heater should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient
(EFT)
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
Not applicable for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line-to-line
± 2 kV line-to-ground
± 1 kV line-to-line
± 2 kV line-to-ground
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0% UT for 0.5 cycle at
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT for 1 cycle
and
70% for 0.5 seconds
0% UT for 5 seconds
0% UT for 0.5 cycle at
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT for 1 cycle
and
70% for 0.5 seconds
0% UT for 5 seconds
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the Test Tube Heater
requires continued operation
during power mains
interruptions, it is
recommended that the Test
Tube Heater be powered from
an uninterruptible power
supply or a battery.
Power frequency
(50/60Hz) magnetic
field IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
Summary of Contents for K-FTH-1012
Page 113: ...K FTH 1012 7 3 K FTH 1012 CE 2012 19 K APS 300 CE 2012 19 IP64 EC REP REF SN 7...
Page 114: ...K FTH 1012 7 4 1 7...
Page 115: ...K FTH 1012 7 5 1 1 36 9 C K FTH 1012 Cook 7...
Page 117: ...K FTH 1012 7 7 3 3 100 240 VAC 50 60 Hz 3 4 EMC IEC 60601 1 2 2014 EMC EMI Cook 7 7...
Page 118: ...K FTH 1012 7 8 4 1 1 K FTH 1012 CP 2 3 4 K FTH 1012 5 4 2 1 2 3 4 5 4 7...
Page 120: ...K FTH 1012 7 10 5 4 5 5 6 6 1 70 6 2 4 3 1 2 3 6 5 mL 4 5 6 7...
Page 121: ...K FTH 1012 7 11 1 2 3 6 5 mL 1 2 3 1 0 C 36 9 C 6 3 SMM30004 SMM30004 6 4 7...
Page 204: ...K FTH 1012 12 4 1 12...
Page 205: ...K FTH 1012 12 5 1 1 36 9 C K FTH 1012 Cook 12...
Page 207: ...K FTH 1012 12 7 3 3 100 240 50 60 3 4 EMC IEC 60601 1 2 2014 EMC EMI Cook 7 12...
Page 208: ...K FTH 1012 12 8 4 1 1 K FTH 1012 CP 2 3 4 K FTH 1012 5 4 2 1 2 3 4 5 4 12...
Page 210: ...K FTH 1012 12 10 5 4 5 5 6 6 1 70 6 2 4 3 1 2 3 6 5 4 5 6 12...
Page 211: ...K FTH 1012 12 11 1 2 3 6 5 1 2 3 1 0 C 36 9 C 6 3 SMM30004 SMM30004 6 4 12...
Page 291: ...Notes...
Page 292: ...Notes...
Page 293: ...Notes...