Technical alarms and status messages
CNAP
®
Monitor 500
Page 14-3
Version 5.2.5,
h
21-FHZU-10002
Message
Priority
Possible cause
Measures
switched off
-
Current measurement discon-
tinued, monitor switched off
automatically
MU: Memory
Full – Dis-
charge Patient
Low *
-
Internal memory is full (as
a result of long measuring
periods without discharg-
ing)
-
Discharge patient via menu
Patient Data to free memory
-
Start new measurement
14.2
BP Wave Out (IBP)
14.2.1
Status messages
Message
Possible cause
Measures
IBP: Connected
-
BP Wave Out
is connected
to patient monitor
-
Perform zeroing (refer to chap-
ter 4.4)
-
Make sure to disable zeroing
when calibration is complete
IBP:
Disconnected
-
BP Wave Out
is discon-
nected from patient monitor
-
n.a.
14.2.2
Technical alarms
Message
Priority
Possible cause
Measures
IBP: Fault
Low *
-
Internal controller prob-
lem
-
Reboot CNAP
®
Monitor 500
-
In case of persistent error,
contact service
IBP:
Transmission
Fault
Low *
-
Interface problem
-
In case of persistent error,
contact service
IBP:
EEPROM RW Error
Low *
-
I/O memory chip defec-
tive
-
Reboot CNAP
®
Monitor 500
-
In case of persistent error,
contact service
IBP:
Iso Board Fault
Low *
-
Isolation board failure
-
In case of persistent error,
contact service
IBP: Iso Board
Bridge Voltage
Low *
-
Bridge voltage > 10V
(BP Wave Out)
-
Disconnect CNAP
®
trans-
ducer cable
-
Check bridge voltage range
of patient monitor (refer to
chapter 4.4)
-
In case of persistent error,
contact service